Diamond UV Detector with Graphite Electrodes Formed by Laser Ablation
A Diamond being a wide-bandgap radiation-hard semiconductor, is rather promising for the development of UV detectors, including the solar-blind ones. A version of an all-carbon UV detector has been implemented on a single-crystal CVD diamond wafer with the interdigital electrodes consisting of a system of shallow (near-surface) and deep graphitized grooves produced by laser ablation without any lithography. Simulation of the electric field distribution in the detector has demonstrated the advantage of using buried electrodes that provide a more uniform field distribution within the crystal. The photosensitivity spectra exhibit a sharp peak at 220 nm, with the buried electrodes providing a 75% higher response amplitude (14 A/W).
V. G. Ralchenko 1, 4, V. P. Pashinin 1, A. F. Popovich 1, 2, A. P. Bolshakov 1, A. K. Martyanov 1, K. V. Zakharchenko 1, 3, V. I. Konov 1, B. Dai 4
Prokhorov General Physics Institute of the RAS, Moscow, Russia
Kotel’nikov Institute of Radio Engineering and Electronics of the RAS, Fryazino, Moscow Region, Russia
Moscow State Technical University of Civil Aviation, Moscow, Russia
Harbin Institute of Technology, Harbin, PR China
A Diamond being a wide-bandgap radiation-hard semiconductor, is rather promising for the development of UV detectors, including the solar-blind ones. A version of an all-carbon UV detector has been implemented on a single-crystal CVD diamond wafer with the interdigital electrodes consisting of a system of shallow (near-surface) and deep graphitized grooves produced by laser ablation without any lithography. Simulation of the electric field distribution in the detector has demonstrated the advantage of using buried electrodes that provide a more uniform field distribution within the crystal. The photosensitivity spectra exhibit a sharp peak at 220 nm, with the buried electrodes providing a 75% higher response amplitude (14 A/W).
Keywords: diamond; UV detector; laser ablation; graphitization
Article received: 28.10.2025
Article accepted:12.11.2025
INTRODUCTION
A Diamond being a wide-bandgap radiation-hard semiconductor (with the bandgap energy of 5.47 eV and a fundamental absorption edge of 225 nm), has high electron and hole mobility, high breakdown voltage, thermal conductivity, hardness and chemical resistance that makes it attractive for the development of high energy particle detectors [1–3], X-ray detectors [4, 5] and UV detectors [6–8]. The UV detectors are of interest for the applications in the field of environmental monitoring, pulsed UV laser radiation control, lithography, and satellite communications. The diamond demonstrates the narrow-band photosensitivity near the band gap edge. The optical absorption coefficient is α = 103 cm‑1 at a wavelength of λ = 220 nm, while decreasing to about 1 cm‑1 for λ = 235 nm [9]. The absorption depth α‑1 for photons at the wavelengths of 220 nm and 225 nm is about 10 μm and 30 μm, respectively, and then it increases sharply for photons with the longer wavelengths. In detectors of the photoresistor type with undoped diamond, the collection of charges generated by the photon absorption is typically performed using the interdigital thin-film metallized electrodes (ITMEs) located on the surface of a diamond plate or film [10–12]. Recently, the ITME options with graphite electrodes that have ohmic contact with the diamond, have been developed [13]. They can be obtained by transforming the diamond surface layer into graphite by the high-temperature heating in a vacuum [13], ion implantation [14], exposure to a catalytically active metal [15] or laser ablation [8, 13, 16]. As in the case of metallic ITMEs, the applied field in the interelectrode gap is significantly non-uniform; the charge carriers are moved along the elongated curved paths, increasing the probability of charge capture by the traps (defects) while reducing the detector efficiency. Another loss factor is related to the fact that a significant part of carriers drifts along the diamond surface in a damaged layer with the thickness of approximately 1 μm [17] that occurs as a result of diamond polishing.
To sidestep such problems, the profiled ITMEs have been proposed in the form of channels with metallized walls, between which a more uniform field distribution is achieved with penetration to a greater depth [18]. The enhanced photoresponse in such a geometry has been experimentally confirmed [19] (in another version, the 3D electrodes in the form of a periodic network of vertical graphite whiskers are generated by laser focusing in the bulk diamond [20, 21]). In this paper, we have combined two approaches, namely the laser drawing of graphite electrodes and their deepening into a diamond substrate that has made it possible to significantly improve the UV detector specifications.
EXPERIMENTAL PROCEDURE
Preparation of electrodes
The detector structures have been generated on the surface of a transparent polished plate made of single-crystal diamond with the dimensions of 8×8×0.4 mm3 synthesized in the microwave plasma. The interdigital electrodes have represented a system of parallel tracks made of graphite-like conductive carbon with the length of 3.0 mm, the width of 10 μm and a distance of 50 μm, formed by the laser-pulse ablation of the diamond. The thickness of the near-surface diamond layer, transformed into a graphite-like material by pulsed heating, has been approximately 300 nm [22]. The radiation has been performed with a scanning beam of a Huaray Cypress‑355–25 laser (with a wavelength of 355 nm, pulse duration of 20 ns, pulse repetition rate of up to 100 kHz, and a maximum power of 25 W). The laser radiation with a pulse energy of 11 µJ has been focused into a spot with a diameter of approximately 10 μm (at the 1/e2 level) on the diamond substrate surface using an aspherical quartz lens with a focal length of 63 mm. The substrate has been placed on a translation stage and moved in the XY plane with a step of 0.5 μm.
To increase the diamond surface absorption in the first pulses and reduce the ablation threshold energies, a copper layer with the thickness of 150 nm has been vacuum-deposited onto the substrate. The shallow (near-surface) graphite electrodes have been formed at a pulse repetition rate of 100 Hz and a scanning speed of 0.5 mm/s, using two beam passes along the path. The deeper graphite grooves have been obtained at a pulse repetition rate of 1 kHz, with all other parameters remaining unchanged. The depth of the near-surface and buried electrodes has been measured using a ZYGO NewView 5000 optical profilometer with the values of 0.7 and 16 µm, respectively. Two ITME arrays, each measuring 2×3 mm² with 20 groove pairs, hereinafter referred to as D0.7 and D16, have been placed on a single substrate to accurately compare their detection properties. On both sides of the array, the electrodes are connected to the continuous graphite termination pads with the dimensions of 2.0×0.8 mm² that have also been formed by laser irradiation (Fig. 1).
Photoresponse measurements
The spectral characteristics have been measured in the wavelength range of 190–400 nm using a Specord M‑400 UV-visible spectrometer, with a D2E deuterium lamp serving as the light source. The spectrometer’s output power has been calibrated using the response of a Hamamatsu S1226–44BQ silicon PiN photodiode with a known spectral characteristic and the dimensions of a photosensitive area of 3.6×3.6 mm2.
Simulation of electric field
The electric field distribution in the detector has been simulated using the ANSYS Electronics Desktop software with comparison of the ITME arrays with a depth of 0, 10, and 20 μm and a distance of 50 μm between the electrodes, i. e., rather close in geometry to the experimental conditions. The field distributions E in the detectors with the ITMEs being near-surface and buried up to 20 μm are shown in Fig. 2. In the case of buried electrodes, the field between them becomes significantly more uniform, and the area of strong field penetrates to a depth close to the electrode depth. The charge carriers generated by photon absorption in the gap between the electrodes move along rather shorter, nearly straight-line paths that reduces the probability of capture on the defects. The photons with energies near the band gap edge (220 nm) are absorbed in a layer with the thickness of approximately 10 μm, while the field in this layer is highly uniform. Therefore, an increased sensitivity of detectors with a three-dimensional geometry is expected.
RESULTS
The graphite provides ohmic contact, as follows from the almost linear current-voltage characteristics (I-V) of the dark current (Fig. 3) that is consistent with the results of other authors [13]. The dark current at a bias of U = 10 V is 4 and 6 pA for the near-surface and buried electrodes, respectively. The photocurrent for them reaches the values of 15 and 24 nA which is more than three orders of magnitude higher than the dark current, and the deep electrodes have a sensitivity 60% higher than the flat ones.
The spectral sensitivity of the produced UV detectors in the wavelength range of 195–400 nm at a voltage of 10 V (field E = 0.2 V/μm) is shown in Fig. 4. The photosensitivity R of deep electrodes reaches 14.3 A/W at a wavelength of 224 nm that is 75% higher than the maximum R for the shallow electrodes (8 A/W). The photocurrent drops sharply towards the longer wavelengths up to 1 A/W for λ = 230 nm. In the visible region (λ = 430 nm) the signal is lower by three orders of magnitude, i. e. the detector can be classified as a “solar-blind” one. The obtained photosensitivity is provided at the level of the known best results for UV detectors on the undoped diamond, both with metallic and graphite ITMEs: 0.1 A/W [19], 9.9 A/W [23], 13.0 A/W [12], 21.8 A/W [8]. In the peak point, photocurrent from the detector D16 exceeds by a factor of 50 the signal from photodiode S1226-44BQ in spite of 2 times lesser photosensitive area of the former.
Photoresponse in the visible area is determined by the optical absorption at the defect and impurity levels within the band gap. We attribute the photocurrent increase for the structure with deep electrodes to more complete collection of the charge carriers generated by absorption at defects in the layers far from the surface. Nevertheless, the spectral selectivity, namely the I(220)/I(400) response ratio, is as high as 103.
CONCLUSION
We have proposed and implemented the design of a UV detector on the undoped single-crystal CVD diamond with interdigital electrodes in the form of a system of grooves deepened into the substrate volume, formed by laser ablation in the direct laser drawing mode. The electrical conductivity of the electrodes is ensured by automatic graphitization of the groove walls during the ablation process. The laser microprocessing allows for rapid and convenient diamond microstructuring without the use of lithography. It has been shown that deepening the electrodes leads to a significant increase in the photoresponse. The obtained photosensitivity is 14 A/W in the wavelength range of 220 nm.
FUNDING
This paper has been supported by the Russian Science Foundation, grant No. 23-42-00120 and the National Natural Science Foundation of China, grant No. 52261135545.
AUTHORS
Viktor G. Ralchenko, Cand.of Sc.(Phys.&Math.), Head of the Diamond Materials Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS; e-mail: vg_ralchenko@mail.ru, Moscow, Russia.
ORCID: 0000-0002-2213-4224
Vladimir P. Pashinin; Researcher, Surface Laser Optics Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS; e-mail: vppashinin@gmail.com, Moscow, Russia.
ORCID: 0000-0002-9257-804X
Alexey F. Popovich; Engineer, Kotel’nikov Institute of Radio Engineering and Electronics of the RAS; Senior Engineer, Diamond Materials Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS; e-mail: lex78@mail.ru; Moscow, Russia.
ORCID: 0000-0002-1389-5697
Andrey P. Bolshakov, Cand.of Sc.(Phys.&Math.), Senior Researcher, Diamond Materials Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS; e-mail: bolshak3@yandex.ru, Moscow, Russia.
ORCID: 0000-0002-9190-5766
Artem K. Martyanov, Cand.of Sc.(Phys.&Math.), Senior Researcher, Diamond Materials Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS; e-mail: art.martyanov@gmail.com, Moscow, Russia.
ORCID: 0000-0002-3196-5695
Kirill V. Zakharchenko, Cand.of Sc.(Phys.&Math.), Researcher, Diamond Materials Laboratory, Natural Sciences Center of Prokhorov General Physics Institute RAS, Associate Professor, Department of Physics, Moscow State Technical University of Civil Aviation; e-mail: nanophys@mail.ru, Moscow, Russia.
ORCID: 0000-0003-3144-714X
Vitaly I. Konov, Dr.of Sc.(Phys.&Math.), acad. RAS, Head of the Natural Scienses Center of Prokhorov General Physics Institute RAS; e-mail: vik@nsc.gpi.ru, Moscow, Russia.
ORCID: 0000-0002-1135-5566
Bing Dai, Cand.of Sc.(Phys.&Math.), Assistant Prof., Harbin Institute of Technology; e-mail: daibinghit@vip.126.com, Harbin, PR China.
ORCID: 0000-0002-1170-4084
CONFLICT OF INTEREST
The authors declare that there is no conflict of interest. All authors have carried out work in the field of their competencies, the manuscript is submitted after a joint discussion and approval of the final version of the text.
rus